PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual
20.2 Create new recipe
1. Select
Create new recipe
in the Recipe MODE tab.
2. Enter the volume, density and viscosity when prompted. All additional
parameters can be adjusted after the recipe is saved.
3. The following parameters are then automatically suggested:
a. Tube size
b. Speed
c. Acceleration
d. Deceleration
e. Anti-drip
4. Enter the first fill delay and between fill delay.
5. Save the recipe—New recipe or overwrite existing.
6. Optional: Edit the all recipe parameters as desired. The recipe name at the top
of the screen will be displayed in red to indicate the recipe has been modified.
7. Save the recipe to confirm changes and the recipe name will be displayed in
white. Refer to "Save recipe" on page 66.
i
Besides the parameters included in a recipe, filling
accuracy is also influenced by fluid path layout. Refer to
"Fluid path layout" on page 23.
20.3 Volume
This sets the target fill amount. To change between volume or weight, refer to
"Units" on page 105.
•
Minimum = 0.0100 millilitres or grams
•
Maximum = 9999.9 millilitres or grams
20.4 Density
Equates the fill mass (weight) measured by a balance to the dispensed fill volume by
the pump.
20.5 Tube size
Defines the default value used for the first calibration and the limits applied to a
calibration.
i
Only compatible Accusil tubing sizes are listed. Refer to
"Tube selection" on page 21.