1
2
C
LE
A
NING S
A
NITIS
A
TION DISINFE
C
TION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket
.
DEVI
C
E
A
ND TU
B
ING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort)
.
ACC
ESSORIES
Open the nebuliser by turning the upper part (
C
1
.3
) anticlockwise, remove the nozzle
(
C
1
.2
) from the upper part (
C
1
.3
) by pressing as shown by the
2
arrows in the "
C
onnection
diagram" in point
C
1
.
Then proceed according to the following instructions
.
S
A
NITIS
A
TION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below
.
method
A
:
Sanitise accessories
C
1.1-
C
1.2-
C
1.3-
C
1.4-
C
2-
C
3-
C
4-
C
5-
C
6-
C
7
uunder
potable hot water (approximately
4
0°
C
) with a gentle, non abrasive dish detergent
.
method
B
:
Sanitise accessories
C
1.1-
C
1.2-
C
1.3-
C
1.4-
C
2-
C
3-
C
4-
C
5-
C
6-
C
7
in the
dishwasher with a hot cycle
.
method
C
:
Sanitise accessories
C
1.1-
C
1.2-
C
1.3-
C
1.4-
C
2-
C
3-
C
4-
C
5-
C
6-
C
7
by immersing
them in a solution of
5
0% water and
5
0% white vinegar, then rinse thoroughly under
potable hot water (approximately
4
0 °
C
)
.
If you want to also perform the cleaning for DISINFE
C
TION, jump to the DISINFE
C
TION
paragraph
.
A
fter having sanitised the accessories, shake them vigorously and place them on a paper
towel
.
A
lternatively, dry them with a jet of hot air (for example, a hair dryer)
.
DISINFE
C
TION
A
fter sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below
.
method
A
:
A
ccessories
C
1.1-
C
1.2-
C
1.3-
C
1.4-
C
2-
C
3-
C
4-
C
5-
C
6-
C
7
can be disinfected
.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy
.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant
.
-
C
ompletely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts
.
Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution
.
USE OF NE
B
ULIZ
A
TION M
A
NU
A
L
C
ONTROL
To achieve continuous nebulisation action you should not use the manual nebulisation
control (
C
7), especially in the case of children or persons with reduced physical, sensory, or
mental capabilities
.
The manual nebulisation control is useful for limiting dispersion of the
medication in the surrounding environment
.
To start nebulizing close
with a finger the hole of the
nebulizer manual control
(
C
7) and breathe in gently;
we recommend to hold your
breath for a moment so that
the inhaled aerosol droplets
can be deposited,
m e a n w h i l e , t o d i s a b l e
nebulizing, remove your fin-
ger from the hole of the
nebulizer manual control to
avoid the waste of drug, op-
timizing its acquisition
.
Then
exhale slowly
.