The device has a suction filter (
A3
) that should be replaced
when it is dirty or changes color
.
Do not wash or reuse the same
filter
.
Regular replacement of the filter is required to help and
ensure correct performance of the compressor
.
C
heck the filter
on a regular basis
.
Do not replace the filter during use
.
Only use original accessories and spare parts by Flaem, we
disclaim any liability in the event of using non original spare parts or accessories.
A
V
A
IL
AB
LE SP
A
RE P
A
RTS
Description
C
ode
- RF7 Dual Speed Nebulizer Set composed of: nebulizer and mouthpiece
- Mask Set composed of: adult mask, pediatric mask and elastic band
- 1 m connection tube
- Spare
A
ir fi lter kit per unit (
2
pcs)
- Nose piece Set composed of: adult nose piece and pediatric nose piece
AC
O
46
1P
AC
O
462
P
AC
O
35
AC
O1
64
P
AC
O
463
P
- Remove the disinfected parts and rinse abundantly with warm potable water
.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer
.
method
B
:
Sanitise the accessories
C
1.1-
C
1.2-
C
1.3-
C
1.4-
C
2-
C
3-
C
4-
C
5-
C
6-
C
7
by boiling
them in water for 10 minutes; use demineralised or distilled water to prevent calcium
deposits
.
method
C
:
Sanitise the accessories
C
1.1-
C
1.2-
C
1.3-
C
1.4-
C
2-
C
3-
C
4-
C
5-
C
6-
C
7
with a hot
steam steriliser for baby-bottle (not the microwave type)
.
Perform the process faithfully
following the instructions of the steriliser
.
To ensure that the disinfection is effective,
choose a steriliser with an operating cycle of at least
6
minutes
.
A
fter having disinfected the accessories, shake them vigorously and place them on a
paper towel
.
A
lternatively, dry them with a jet of hot air (for example, from a hair dryer)
.
A
t the end of each use store the device complete with accessories in a dry place away
from dust.