Automatic Cleaning and Disinfection
It is recommended to use a
thermo-disinfector to clean and disinfect the Lip
Hook, File Clip and Contra Angle Holder.
Put the Lip Hook, File Clip and Contra Angle Holder on the tray of the
thermo-disinfector, select the "surgical device", and start the automatic cleaning
and disinfection procedure.
Cleaning and disinfection procedure:
a) Pre-cleaning: Pre-clean with tap water less than 40 ° C for 4 minutes
b) Cleaning: Immerse and wash with a multi-enzyme cleaner for 6 minutes at 55 °C
c) Rinse stage I: Flush with tap water less than 40 ° C for 1 minute
d) Rinse stage II: Flush with tap water less than 40 ° C for 1 minute
e) Disinfection: The temperature is 80 °C and the action time is 10 min
f) Drying: The temperature is 100 °C and the action time is 15 min
Caution
⚫
The user must follow the special instructions of the manufacturer of the fully
automatic washing machine. In order to ensure the cleaning and disinfection
effect, the cleaning and disinfection time should not be less than the time
recommended by the manufacturer.
⚫
We recommend the use of proven LIRCON® multi-enzyme cleaning solution or
multi-enzyme cleaning solution that complies with local regulations (e.g. CE,
FDA approval).
⚫
Please use a thermo-disinfector that meets the requirements of ISO 15883.
⚫
Considering that some countries have different requirements for A0 values,
please refer to ISO 15883 for temperature and time of disinfection.