32
7.3
Manual Cleaning
a)
Rinsing the Lip Hook, File Clip and Contra Angle Holder in flowing tap water
(<40°C).
b)
After cleaning, remove any liquid residue with a lint-free cotton cloth and then
dry with compressed air (1-2 Bar).
Warning
Do not put the Lip Hook, File Clip, and Contra Angle Holder into the
disinfectant container for immersion and disinfection. If the internal liquid is not
cleaned and the drying is not complete, the internal parts will be corroded.
7.4
Manual Disinfection
Place the lint-free cotton cloth in a container filled with LIRCON® medical device
cleaning solution (metal type), wring it out and wipe the Test Wire, Lip Hoop, File
Clip, Contra Angle Holder/Lighting device and Contra Angle at least 3 times.
It is recommended to use the medical device cleaning solution (metal type) of
LIRCON®, which with an effective chlorine content of 1.0%-1.2% (W/V).
Warning
Do not put the Test Wire, Lip Hoop, File Clip, Contra Angle Holder/Lighting
device and Contra Angle into the disinfectant container for immersion and
disinfection. If the internal liquid is not cleaned and the drying is not complete, the
internal parts will be corroded.
Caution
If you want to use other disinfectant, use a disinfectant that complies with
local national regulations (such as CE certification, FDA certification), and follow
the instructions provided by the disinfectant manufacturer.