31
7.
Cleaning ,Disinfection and Sterilization
Note
Cleaning, disinfection, and sterilization have limited impact on the reusable parts
of the motor handpiece. Therefore, the number of times the procedure is repeated
is determined by the degree of wear of the part. If visual inspection reveals
damaged parts, stop using them and purchase new parts from the manufacturer or
dealer.
AUTION:
•
Refer to the contra-angle operation manual for all operations regrading
the contra-angle.
7.1
Preparation after Using
Immediately after use, the reusable parts should be immersed in tap water <40 ° C
(The quality of drinking water, the ‘water’ mentioned in this chapter, is required to
meet this standard.) to remove dirt. Do not use a fixed detergent or warm water
(>40 °C), as this will cause the residue to be fixed and affect the post-treatment
effect.
Transportation: Transport to the post-processing area for safe storage to avoid any
damage and environmental pollution.
7.2
Preparation before Cleaning
7.2.1
Disassemble the Lip Hook, File Clip and Contra Angle Holder and place them
in a stainless steel box.
7.2.2
Decontamination preparation (Pre-cleaning)
Flush the Lip Hook, File Clip and Contra Angle Holder with running tap water <40
°C until all visible residue is removed.