Troubleshooting
48
Error
Symptom
Possible causes and corrective measures
E-22/E-23
Rotor stops without
deceleration to standstill.
Instrument cannot be
operated.
Error in speed entry
Switch the instrument off and on again.
If the error persists, contact service representative.
E-24
Instrument cannot be
operated.
Wrong status information from the lid-latch.
1. Turn the centrifuge off and on.
2. After re-switching on, the display shows "
Lid FAiL
"
3. If the centrifuge lid has been already opened, the display
shows "
Close Lid
". Thereupon close the lid.
4. The centrifuge tries to open the lid to switch for starting
the normal operation mode.
If the error persists, contact service representative.
E-29
Motor does not start.
Motor or rotor blocked.
1. Switch instrument off and on again using the mains
switch.
2. Open the centrifuge lid.
3. Check whether the rotor can turn freely.
If you cannot clear the malfunction, contact a service
representative.
Summary of Contents for AccuSpin Micro 17
Page 1: ...Instructions for use AccuSpin Micro 17 17R ...
Page 10: ...For your safety 6 for your notes ...
Page 46: ...Maintenance and care 42 ...
Page 56: ...Troubleshooting 52 for your notes ...
Page 62: ...Technical data 58 for your notes ...
Page 63: ...Appendix 59 Appendix ...
Page 69: ...Appendix 65 For your notes ...
Page 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Page 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Page 72: ...Appendix 68 For your notes ...
Page 78: ...74 for your notes ...