![Fisher Scientific AccuSpin Micro 17 Instructions For Use Manual Download Page 16](http://html.mh-extra.com/html/fisher-scientific/accuspin-micro-17/accuspin-micro-17_instructions-for-use-manual_2296621016.webp)
Accessories
12
Rotors for the accuSpin Micro 17
Table 1: Performance characteristics for accuspin Micro 17
rotor designation
order no.
Micro liter-rotor
24 x 2 ml
75003524
Micro liter-rotor
36 x 0,5 ml
75003436
Dual-rotor
18 x 2 ml / 0,5 ml
75003418
places / volume
24 x 1.5 / 2 ml
36 x 0.5 ml
18 x 2 ml + 18 x 0.5 ml
maximum permissible load [ g ]
24 x 4
36 x 0.5
18 x 4 + 18 x 0.5
minimum speed n
min
[ min
-1
]
300
300
300
maximum speed n
max
[ min
-1
]
13 300
13 300
13 300
maximum RCF value at n
max
17 000
15 600
16 800
acceleration / deceleration time [ s ]
11/12
9/10
11/12
radius maximum/ minimum [ cm ]
8.6 / 5.1
7.9 / 5.0
8.5 / 4.8
angle [ ° ]
45
45
45
heating of samples at n
max
[ °C ]
relative to room temperature 23°C,
run time 1 hour
33 31 33
aerosol-tight *
yes
no
no
permissible temperature range
autoclavable (number of cycles)
-4 °C to +40 °C
121°C, (20 cycles)
-4 °C to +40 °C
121°C, (20 cycles)
-4 °C to +40 °C
121°C, (20 cycles)
* Tested and approved by HPA, Porton-Down, UK – Please see the references in the chapter
"Aerosoltight application"!
Summary of Contents for AccuSpin Micro 17
Page 1: ...Instructions for use AccuSpin Micro 17 17R ...
Page 10: ...For your safety 6 for your notes ...
Page 46: ...Maintenance and care 42 ...
Page 56: ...Troubleshooting 52 for your notes ...
Page 62: ...Technical data 58 for your notes ...
Page 63: ...Appendix 59 Appendix ...
Page 69: ...Appendix 65 For your notes ...
Page 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Page 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Page 72: ...Appendix 68 For your notes ...
Page 78: ...74 for your notes ...