Troubleshooting
43
Troubleshooting
Mechanical emergency lid release
In case of a power failure you cannot open the lid nor-
mally using the electric emergency release. To permit
unloading even in this case, the centrifuge is equipped
with a hand operatated lid unlocking system. However,
you may use this system
only
in case of emergency.
Rotor can spin at high speed!
Touching it may cause severe inju-
ries!
Always wait for several minutes until
the rotor has come to a complete
standstill without deceleration. With-
out power the brake does not func-
tion, and braking takes much longer
than normal!
Should it be necessary to open the lid manually, pro-
ceed as follows using an appropriate/special tool:
1. Make sure the rotor stands still. (control window).
2. Unplug the mains cord.
3. Push a small straight wire of a length of approx.
7cm (e.g. a bended paper clip etc.) through the hole
located in the middle above the display, at the
chamber edge.