Handling the rotor
17
Handling the rotor
Rotor temperature
The rotors are only to be used within
the temperature range from - 9°C to
+40°C. Pre-tempering in a freezer
below - 9°C is vorbidden.
Lifetime of the rotor
There is no limitation on the service life of the high
performance rotors. However please observe the fol-
lowing due to safety reasons:
Rotors and accessories made of
plastic should not be exposed to
direct sunlight and UV radiations
.
If the rotor shows signs of discolor-
ation, deformation or wear, or imbal-
ance it has to be exchanged straight
away!
Rotor lid with a spring lock
Opening
The rotor lid is held on the central rotor nut integrated
into the rotor
The rotor lid is opened by pressing the red unlocking-
button in the central handhold of the rotor lid.
The lid can be lift-off easy.
Handhold
Unlocking-button
Summary of Contents for AccuSpin Micro 17
Page 1: ...Instructions for use AccuSpin Micro 17 17R ...
Page 10: ...For your safety 6 for your notes ...
Page 46: ...Maintenance and care 42 ...
Page 56: ...Troubleshooting 52 for your notes ...
Page 62: ...Technical data 58 for your notes ...
Page 63: ...Appendix 59 Appendix ...
Page 69: ...Appendix 65 For your notes ...
Page 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Page 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Page 72: ...Appendix 68 For your notes ...
Page 78: ...74 for your notes ...