Troubleshooting
46
Error
Symptom
Possible causes and corrective measures
–
Exceptionally running
noise.
Imbalance.
1. Stop the centrifuge by pressing the "stop" key, in case of
emergency, unplug mains power cord.
2. Wait until the centrifuge comes to a complete stop.
3. Check whether the rotor is properly loaded.
4. Check whether a broken tube, damage to the rotor or
motor is responsible for the run noise.
If you cannot locate and solve the problem yourself, contact a
service representative
Display "oP" appears
although lid is closed
.
Will not start.
Centrifuge lid not properly closed
- Open the lid and repeat locking procedure.
If you cannot locate and solve the problem yourself, contact a
service representative.
“Lid” Rotor
stops
with
deceleration to standstill.
Centrifuge lid was opened manually during the run.
- Close centrifuge lid immediately
The instrument phases out with deceleration.
For further centrifugation, you have to switch the instrument
off and switch it on again.
Summary of Contents for AccuSpin Micro 17
Page 1: ...Instructions for use AccuSpin Micro 17 17R ...
Page 10: ...For your safety 6 for your notes ...
Page 46: ...Maintenance and care 42 ...
Page 56: ...Troubleshooting 52 for your notes ...
Page 62: ...Technical data 58 for your notes ...
Page 63: ...Appendix 59 Appendix ...
Page 69: ...Appendix 65 For your notes ...
Page 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Page 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Page 72: ...Appendix 68 For your notes ...
Page 78: ...74 for your notes ...