Maintenance and care
40
5. Treat the rotor and the rotor lid according to the
instructions given for the disinfectant (soaking in
liquid or spraying). You must strictly observe the
specified action times!
6. Turn the rotor head down and drain off the disinfec-
tant. Thereafter thoroughly rinse rotor and lid with
water.
7. Dispose of the disinfectant according to valid regu-
lations.
8. Aluminum rotors have to be treated with anticorro-
sive protective oil subsequently.
9. All seals have to be lubricated again.
Disinfection with bleaching lye
These agents contain highly ag-
gressive hypochlorites and must
not be used with aluminum rotors!
The following precautionary measures are to be taken
for extensive protection of the plastic rotors:
1. Avoid high temperatures!
The bleaching solution and the rotor should not be
warmer than ca. 25 °C.
2. Do not let the bleaching solution act longer than
absolutely necessary!
3. After disinfection, rinse the rotor thoroughly with
distilled water and allow drying.
4. All seals have to be lubricated again
Summary of Contents for AccuSpin Micro 17
Page 1: ...Instructions for use AccuSpin Micro 17 17R ...
Page 10: ...For your safety 6 for your notes ...
Page 46: ...Maintenance and care 42 ...
Page 56: ...Troubleshooting 52 for your notes ...
Page 62: ...Technical data 58 for your notes ...
Page 63: ...Appendix 59 Appendix ...
Page 69: ...Appendix 65 For your notes ...
Page 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Page 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Page 72: ...Appendix 68 For your notes ...
Page 78: ...74 for your notes ...