FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
getYOFFSET##
- Return the y Chromaticity offset
Transmit
Receive
getyoffset##
±
0.yyy
Where:
##
represents the Fiber Number and is a number in the range 01 – 20.
±0.yyy
represents the y Chromaticity offset
Description
This command is used to return the y Chromaticity offset for the LED under the Fiber number
##. The value of this offset must be set by the
setyoffset##
command. The default value
is 0.000.
Example:
The PC transmits
getyoffset01
to the LED Analyser to instruct it to send the stored y
Chromaticity offset for the LED positioned under Fiber No 1. The LED Analyser will return a
string
±0.yyy
to the PC.
getyoffset01
-0.025
63