FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getWI##
- Get the Dominant Wavelength & Intensity
Transmit
Receive
getwi##
getwiall
XXX YYYYY
Where:
##
represents the Fiber Number and is a number in the range 1 – 20.
all
represents the 20 Fibers in a Feasa 20F, or 3 Fibers in a 3F.
XXX
represents the dominant wavelength of the LED in nanometers.
YYYYY
represents the intensity of the LED.
Description
This command is used to get the value of the Dominant Wavelength for the LED under the
Fiber number
##
.This command should be preceded by the
capture C
command to ensure
valid LED data is stored in the memory of the LED Analyser. The
C1-C5
commands are
available but are not as accurate. The data for each LED can be read out one-at-the-time, in
any order.
Alternatively use the
getwiall
command to read back the LED data for all 20 leds in
one command.
Example:
The PC transmits
getwi01
to the LED Analyser to instruct it to send the stored dominant
wavelength & Intensity for the LED positioned under Fiber No 1. The LED Analyser will return
a string
xxx yyyyy
to the PC.
getwi01
513 12345
32