![Feasa 20-F User Manual Download Page 61](http://html1.mh-extra.com/html/feasa/20-f/20-f_user-manual_3778790061.webp)
FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
set/putYOFFSET##0.xxx
- Set the y Chromaticity Offset
Transmit
Receive
setyoffset##
±
0.xxx
putyoffset##
±
0.xxx
OK
OK
Where:
##
represents the Fiber Number and is a number in the range 01 – 20.
0.xxx
represents the y Chromaticity offset value (±0.000 – 0.300).
Description
This set command is used to set an offset to the displayed y Chromaticity value. The limit of
the offset is ±0.300 which means values must be in the range ±0.000 – 0.300. This
command is useful when the user wishes to set the y Chromaticity to be the same as that
specified by the LED Manufacturer. The default value of the offset is 0.000. The offset is
stored in non-volatile memory and will remain at the programmed setting until changed by a
new
Setyoffset
command.
This put command is used to set an offset to the displayed y Chromaticity value. The limit of
the offset is ±0.300 which means values must be in the range ±0.000 – 0.300. This
command is useful when the user wishes to set the y Chromaticity to be the same as that
specified by the LED Manufacturer. The default value of the offset is 0.000. The offset is
stored in RAM memory and will remain at the programmed setting until changed by a new
Setyoffset
command or power OFF.
Example:
The PC transmits
setyo0.050
to the LED Analyser to instruct it to set the y offset on
Fiber 01 to +0.050.
setyo0.050
OK
Example:
The PC transmits
putyo0.050
to the LED Analyser to instruct it to set the y offset on
Fiber 01 to +0.050.
putyo0.050
OK
61
Summary of Contents for 20-F
Page 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Page 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Page 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Page 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...