FEASA LED ANALYSER
FUNCTIONAL VERSION
General Commands
EOT -
End of Transmission
Character
Transmit
Receive
enableeot
disableeot
See details below
Description
Every command sent to the LED Analyser or response received from the Led Analyser, ends
with a set of characters that indicates that the command or response has finished. Those
characters are CR + LF, which are ASCII characters 13 and 10 respectively. Alternatively,
responses from the LED Analyser can also have an end-of-transmission format like CR + LF +
EOT, which are the ASCII characters 13, 10 and 04 respectively. CR + LF are always present,
but character EOT can be enabled or disabled using the commands ENABLEEOT and
DISABLEEOT. Having EOT character enabled can be extremely useful in multi-line commands,
since this allows to differentiate between the end of a line and the end of a transmission.
The EOT command is disabled by default. To enable EOT simply send the command enableeot
using Feasa Terminal software. The Analyser will maintain the EOT state even when powered
off. To disable the state just send the command disableeot.
Information Received back from the Analyser using the
getstatus
command is as follows:
Hardware: Feasa xx-y
Serial number : Xyyy
Firmware Version : Iyyy
Intensity Mode : Logarithmic
Photopic response: ON
Exposure Factor : 01
Capture : Manuel 04
Auto PWM mode : OFF
Number of Fibers : 05
External Trigger status: Disabled
Colour Gains: 100 150 100
End of Transmission Character:
Disabled
Selected calibration set: 0
user calibration date: dd/mm/yyyy
Factory calibration date: dd/mm/yyyy
48
Summary of Contents for 20-F
Page 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Page 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Page 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Page 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...