FEASA LED ANALYSER
FUNCTIONAL VERSION
Daisy Chain Mode
Example
Busfree
'free the bus
OK
busc3
'All Analysers capture using range 3 High
OK
busceF304
'poll the analyser F304
1
busgetF304
'Connect to the Analyser F304
OK
getserial
'Confirm s/n of the Analyser
F304
gethsi01
'Get the Data for Fiber 1
000.51 100 36491
gethsi02
.
.
gethsi20
120.51 100 66542
'Data for Fiber 20
busfree
OK
busceF461
1
busgetF461
'Connect to the next Analyser F461
OK
getserial
'Confirm to Analyser
F461
gethsi01
'Get the Data for Fiber 1
000.51 100 36491
gethsi02
gethsi20
'Get the Data for Fiber 20
.
Repeat this sequence for all Analysers in the chain.