FEASA LED ANALYSER
FUNCTIONAL VERSION
External Trigger Mode
External Trigger Method:
Functional Led Analysers have a 2 pin input port (Tr) to activate an external trigger capture.
Prior to activation the programmer must set the
External Trigger Capture mode
and the
External Trigger Output mode.
When set to the correct modes
connecting
pin 1 to pin 2
will activate a capture and output the relevant readings as selected. No External Voltages are
required.
Transmit
Receive
Setexternalcapture##
Putexternalcapture##
getexternalcapture
See details below
Where
##
is as follows:
00
Disable Trigger
01
Capture1 or C1 (Range Low)
02
Capture2 or C2 (Range Medium)
03
Capture3 or C3 (Range High)
04
Capture4 or C4 (Range Super)
05
Capture5 or C5 (Range Ultra)
06
Capture or C (Range Auto)
10
Capturepwm or CPWM (Range Auto)
11
Capturepwm1 or CPWM1
12
Capturepwm2 or CPWM2
13
Capturepwm3 or CPWM3
14
Capturepwm4 or CPWM4
15
Capturepwm5 or CPWM5
50
Sequencecapture mode
Transmit
Receive
Setexternaloutput####
Putexternaloutput####
getexternaloutput
See details below
Where
####
is as follows:
No output
setexternaloutputnone
HSI
setexternaloutputhsi
XYI
setexternaloutputxyi
RGBI
setexternaloutputrgbi
CCTI
setexternaloutputcct
ABSINT
setexternaloutputabsint
WSI
setexternaloutputwsi
75
Summary of Contents for 20-F
Page 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Page 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Page 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Page 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...