Get Data Commands
The
get data
commands are used to read out the Colour, Saturation and Intensity data stored
by the
capture
commands.
The data from the last
capture
command remains in memory until a new
capture
command is
issued or the power is removed from the Analyser.
Commands are transmitted and received using ASCII characters and are
NOT case-sensitive
.
All commands must be terminated with a
<CR>
or
<LF>
character.
Under Range Condition
An under range condition will occur when insufficient light from the LED reaches the sensor for
the range selected. This will be indicated by
999.99 999 00000
for
HSI,
000 000 000 00000
for
RGBI
and
0.0000 0.0000
for xy and uv.
If this condition occurs select the next
lower
range and test again.
Over Range Condition
An over range condition will occur when too much light from the LED reaches the sensor for
the range selected by the switch. This will be indicated by
999.99 999 99999
for
HSI,
255 255 255 99999
for
RGBI
and
0.0000 0.0000
for xy and uv.
If this condition occurs select the next
higher
range and test again.
Incorrect Capture Mode
An incorrect capture mode condition will occur when a blinking (pwm) light from the LED
reaches the sensor when the Autorange C capture is selected. This will be indicated by
XXX.XX XXX XXXXX
for
HSI, XXX XXX XXX XXXXX
for
RGBI
and
X.XXXX X.XXXX
for xy and uv.
The manual Capture Ranges C1 – C5 are unaffected by this.
If this condition occurs in Software Rev F201 – F206 use the
CPWM
command instead of the
Capture
command.
For software Rev F207 upwards you can set the capture command to mimic the cpwm
command by typing
setautopwm1
. This can be verified using the getstatus or getautopwm
commands. To turn this function off type the command
setautopwm0
.
24
Summary of Contents for 20-F
Page 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Page 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Page 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Page 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...