FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
ResetRGBM -
Reset the RGB usercal setting to factory default
Transmit
Receive
resetrgbm@@
resetrgbmall
OK
Where:
@@ is the individual fiber number to be reset from 01-20
ALL will reset all 20 fibers in a specific calibration set
Description:
This command will reset the RGB calibration values (of the RGB calibration mode) to 0 for a
particular fiber or for all fibers in a Calibration set. The Led analyser is capable of storing fifty
one (00-50) separate user Calibration sets on the board. Each one must be reset separately.
Example:
If you want to reset RGB values for all fibers in calibration set No.25
setcalibration25
OK
resetrgbmall
OK