FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
3Vrms
3V/m
Portable and mobile RFcommunications
equipment should be used no closer to any part
of the monitor, includingcables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d =
1.2
P
d =
1.2
P
80MHz to 800MHz
d =
2.3
P
800MHz to 2.5GHz
where
P
is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is
the
recommended separation distance in meters