II-
10
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
(11) Pulse Rate
Item
Specification
Measurement
range and accuracy
a)
Analog SpO
2
module
Measuring range: 25bpm~250bpm;
Resolution: 1bpm;
Accuracy: ±1bpm.
b) Masimo SpO
2
module
Measurement range: 25bpm~240bpm;
Resolution: 1bpm;
Accuracy: ±3bpm (with motion) or ±5bpm (without motion).
c)
Nellcor SpO
2
module
Measurement range: 25bpm~250bpm;
Resolution: 1bpm;
Accuracy: ±3bpm within the measurement range of 20bpm
~
250bpm;
Not specified within the measurement range of 251bom
~
300bpm.
Alarm range and
accuracy
Alarm range: 20 bpm
~
300 bpm
Accuracy: ±1bpm
(12) Body Temperature
Item
Specification
Measurement
range
0
℃
~50
℃
Accuracy
±0.2
℃
(sensor included)
Alarm range
0
℃~
50.0
℃
(32
℉~
122
℉
)
Alarm resolution
±
0.1
℃
(
±
0.2
℉
)
Display resolution
0.1
C
Operating mode
Derect mode
Transient response
No less than 20 seconds