V-1
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Appendix V EMC
Attention
The monitor meets the requirement of electromagnetic compatibility in IEC60601-1-2.
The user needs to install and use according to electromagnetism compatibility
information which is attached with it.
Portable and mobile RF communicationdevices may influence the monitor performance, so the
monitor should be kept away from them during using.
Guidance and manufacturer’s declaration stated in the appendix.
Warning
The monitor should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the monitor should be observed to verify normal operation
in the configuration in which it will be used.
Class A equipment is intended for use in an industrial environment. The monitor may be
potential difficulties in ensuring electromagnetic compatibility in other environments, due to
conducted as well as radiated disturbances.
Guidance and manufacturer’s declaration –electromagnetic emissions
The monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the monitor should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The monitor is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.