9-1
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Chapter 9 Battery
9.1 General Introduction
The monitor is equipped with a built-in rechargeable batterywhich will be auto charged when the AC power
supply is connected no matterwhether the monitor is turned on or not. In case of sudden power failure, the
monitor will be auto powered by the battery without operation interruption and the
battery workindicator will be lit up 30s after the power failure.
To understand the battery icon:
: indicating the battery is fully charged.
: indicating moderate battery power.
: indicating low battery power.
: indicating critically low battery power, please charge the battery immediately.
: indicating no battery installed or damaged battery.
Attention
Please remove the battery if it will be not be used for a long time and store it properly.
If a battery is installed in the moitor, the battery must be recharged after each use to ensure
enough battery power.
Warning
The battery liquid is harmful, In case the liquid contacts your skin or eye, wash it
immediately with clean water or seek medical advise.
Keep the battery out of the reach of children.
9.2 Low Battery Alarm
The monitor will automatically shut down in case of two low battery. When the battery icon turns to
, indicating the battery power is less than 10%, the monitor will trigger a technical alarm with
alarm message of [ Low Power level! Please charge the battery]. At this moment, the monitor should be
immediately connected to AC supply to charge the battery, or the monitor will auto shut down before the
battery is fully used up.