II-7
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
(8) Respiration
Name
Specification
Method
Thoracic impedance method
Measurement
range
and
accuracy
Range
Adult
6
~
120 rpm
Pediatric/neonate
6
~
150 rpm
Accuracy
7rpm
~
150rpm: ±1rpm or 1%, whichever is the larger;
0rpm
~
6rpm
:
Not specified
Alarm range and
accuracy
Adult
7rpm
~
120rpm
Pediatric/ neonate
7rpm
~
150rpm
Accuracy
±1bpm
Apnea alarm time
10s, 15s, 20s, 25s, 30s, 35s, 40s,error:
±
5s.
(9) NIBP
Name
Specification
Measuring method
Oscillation method
Measurement
range and accuracy
Adult
Systolic pressure
40mmHg
~
270mmHg(5.3kPa
~
36.0kPa)
Mean arterial pressure
20mmHg
~
235mmHg(2.7kPa
~
31.3kPa)
Diastolic pressure
10mmHg
~
215mmHg(1.3kPa
~
28.7kPa)
Pediatric
Systolic pressure
40mmHg
~
200mmHg(5.3kPa
~
26.7kPa)
Mean arterial pressure
20mmHg
~
165mmHg(2.7kPa
~
22.0kPa)
Diastolic pressure
10mmHg
~
150mmHg(1.3kPa
~
20.0kPa)
Neonate
Systolic pressure
40mmHg
~
135mmHg(5.3kPa
~
18.0kPa)
Mean arterial pressure
20mmHg
~
110mmHg(2.7kPa
~
14.7kPa)
Diastolic pressure
10mmHg
~
100mmHg(1.3kPa
~
13.3kPa)
Accuracy
The maximum mean deviation is
±
5mmHg(
±
0.667kPa) and the
maximum standard deviation is 8mmHg(
±
1.067kPa).
Overpressure
protection
Adult
300mmHg
Pediatric
240mmHg
Neonate
150mmHg
Error
±3mmHg
Static
pressure
measurement
range
0mmHg~300mmHg,
Accuracy
±3mmHg
Alarm range and
accuracy
Adult
Systolic pressure:
40mmHg
~
270mmHg(5.3kPa
~
36.0kPa)
Mean arterial pressure:
20mmHg
~
235mmHg(2.7kPa
~
31.3kPa)
Diastolic pressure
10mmHg
~
215mmHg(1.3kPa
~
28.7kPa)
Pediatric
Systolic pressure:
40mmHg
~
200mmHg(5.3kPa
~
26.7kPa)