Chapter 19 TEMP Monitoring
19.1 Overview
The probe should be applied to skin or anus according to the type of the probe you bought. The
recommended minimum measuring time is 1 minute, or the measured value will not be accurate
19.2 TEMP Display
TEMP alarm
limit
Unit
TEMP measured
value
19.3 Safety Information
Warning
Check whether the TEMP probe cable is in good condition prior to monitoring. When you
discennect the TEMP probe cable from the monitor, the monitor will trigger a technical alarm
and display alarm message “T1 SENSOR OFF” on the screen.
Be careful in handling the TEMP probe and cable, when not in use, the probe and cable should
be pulled into the loose ring. If the wire is pulled too tight, it will lead to mechanical damage.
Calibrate the temperature measuring instrument at least once every two years ( or according to
the required time in the hospital directive rules). Please contact the manufacturer when
calibration is needed.
Attention
Disposable temperature probe can only be used once.
During the monitoring process, the temperature measuring instrument will automatically
check itself once per hour. Self-checking will last 2 seconds and will not affect the temperature
monitoring.