2-31
1
2
W
eb Settings Screen and Function Details
Storage Screen
2-9-3 Quota
This is the number of data entries to save until the storage function is stopped when the storage function is in
use.
Input range: 1 to 100,000
Default value: 1000
2-9-4 Rate
Sets the interval during which data should be saved for all measured data.
Not all measurement data is saved. Saving is done intermittently to reduce the amount of data stored and to
facilitate processing.
“0”:
All
obtained
data
is
stored.
Number (positive values only): For each data entry stored, a single measurement value is saved for
HDFKVSHFL¿HGVDPSOLQJ
Input
range:
0
to
1,000,000
Default
value:
10
([)RUH[DPSOHLILVVSHFL¿HGDVWKHLQWHUPLWWHQWVHWWLQJWKHIROORZLQJRSHUDWLRQZLOOEHSHUIRUPHG
Ɣ'DWDIRUZKLFKERWKPHDVXULQJDQGVDYLQJDUHSHUIRUPHG
ż'DWDIRUZKLFKPHDVXULQJLVSHUIRUPHGEXWVDYLQJLVQRW
2-9-5 Start
condition
Sets the “Start trigger” conditions used as the basis for starting storage.
Immediate:
Storage
starts
immediately.
Alarm Off edge: Storage starts from the instant measurement becomes possible.
Alarm On edge: Storage starts from the instant measurement becomes impossible.
Threshold Up edge: Storage starts when the measured value exceeds the set threshold.
Threshold Down edge: Storage starts when the measured value falls below the set threshold.
Threshold Both edge: Storage starts when the measured value exceeds or falls below the set threshold.
Default
value:
Immediate
2-9-6 Threshold
When “Start condition” is “Threshold Up edge,” “Threshold Down edge,” or “Threshold Both edge,” this
VHWWLQJVSHFL¿HVWKHWKUHVKROGWREHXVHGDVWKHUHIHUHQFHIRUWKHVWDUWFRQGLWLRQ
Input range: ±2000.000000 (mm)
Default value: 0