2-14
Measurement Screen
2-5-12
0HGLDQ¿OWHU
*
>&'$PHQX0HGLDQ@
$SSOLHVDPHGLDQ¿OWHUWRSUHYHQWYDULDWLRQE\FXWWLQJVXGGHQFKDQJHVLQWKHPHDVXUHGYDOXH
2)) 1RPHGLDQ¿OWHULVXVHG
$SSOLHVWKHPHGLDQ¿OWHUWKHVSHFL¿HGQXPEHURIWLPHV
Default value: 31
([:LWKDSRLQWPHGLDQ¿OWHU
Number of measurements
1
2
3
4
5
6
7
8
9
10
Measured value
1.9
1.8
2.0
2.1
2.0
2.2
2.5
2.0
1.8
2.1
VW¿OWHUUDQJHDQGUHVXOWV
2.0
QG¿OWHUUDQJHDQGUHVXOWV
2.0
UG¿OWHUUDQJHDQGUHVXOWV
2.0
)RUH[DPSOHZKHQWKHPHDVXUHGYDOXHVIRUWKH¿UVW¿OWHUUDQJHDUHVRUWHGLQDVFHQGLQJRUGHUWKH
values will be as follows.
In this case, the central (No. 4) value (2.0) will be output as the measured value.
Before sorting
1.9
1.8
2.0
2.1
2.0
2.2
2.5
After sorting
1.8
1.9
2.0
2.0
2.1
2.2
2.5
2-5-13 Moving average
Averaging is performed the number of times for the measured value.
7KLVSURFHVVLVDSSOLHGDIWHUPXOWLSO\LQJE\WKHPHGLDQ¿OWHU
Although the measured values are changed gradually, a higher averaging count will take longer for the
measured value to be produced.
1 / 2 / 4 / 8 / 16 / 32 / 64 / 128 / 256 / 512 / 1024 / 2048 / 4096 / 8192 / 16384 /
>8QLW&RXQW@
Default value: 256
Ex.: With an averaging count of 4:
7KHPHDVXUHPHQWUHVXOWVIRUWKHVSHFL¿HGFRXQWZLOOEHDYHUDJHGDQGRXWSXWDVWKHPHDVXUHGYDOXHDV
VKRZQLQWKHIROORZLQJ¿JXUH
Number of measurements
1
2
3
4
5
6
7
8
Measured value
1.9
1.8
2.0
2.1
2.0
2.2
2.5
2.0
VW¿OWHUUDQJHDQGUHVXOWV
1.95
QG¿OWHUUDQJHDQGUHVXOWV
1.975
UG¿OWHUUDQJHDQGUHVXOWV
2.075
WK¿OWHUUDQJHDQGUHVXOWV
2.2
WK¿OWHUUDQJHDQGUHVXOWV
2.175