Software Description
The software adds at the end of a list file a small data block containing the counter values.
The format follow the scheme of scope mode data, it can be distinguished by the waveform length that is
only 12 16-bit words.
The lowest two bits 0 and 1 are the channel number 0..3 for ADC1..ADC4.
Bit 3 indicates scope mode. It is on, so more data words follow.
Bit 4 – 47 are zero.
Bit 48 – 63 are the length of the counter data minus one in units of 16 bit words. This size value is 11 or in
hex format b for 12 16-bit words or 3 64-bit counter data. The counter data follow then next, real time, live
time and event counter. The counter data are 64 bit values
Here is an example of a counter data block:
000b000000000008
0000000007735943
00000000075aacad
0000000000002710
000b000000000009
0000000007735943
00000000075aaed3
0000000000002710
000b00000000000a
0000000007735943
00000000075ab725
0000000000002710
000b00000000000b
0000000007735943
00000000075aaf24
0000000000002710