Installation procedure
In the ADC Settings dialog enable the Trapezoidal Filter (ref. Fig. 2.24).
The resulting scope picture shows the trapezoidal shaped pre-amplifier pulses as in Fig. 2.26. With the now
open Pole Zero Adjust window you may change the settings of the pole zero filter and optimize the resulting
shaped pulse form.
F
ComTec GmbH
25
Fig. 2.24: Enable the Trapezoidal Filter
Fig. 2.26: Scope Picture of the trapezoidal shaped Test Pulser Signals
Fig. 2.25: Effects of different Pole Zero Settings