10
EN
Remove the LARYVOX
®
STYLE HME HighFlow filter cassette with LARYVOX
®
STYLE TOP from LARYVOX
®
STYLE CAP.
To do this, hold the LARYVOX
®
STYLE HME HighFlow filter cassette steady with
two fingers and remove LARYVOX
®
STYLE CAP upward ⑬.
Now press the LARYVOX
®
STYLE HME HighFlow filter cassette together with your
thumb at the predetermined breaking point ⑭ until the filter cassette is released
from LARYVOX
®
STYLE TOP.
IX. SERVICE LIFE
LARYVOX
®
STYLE HME HighFlow filter cassettes
The maximum period of use / wearing period is 24 hours.
The filter must be changed even several times a day if required, for instance if the
respiratory resistance is in danger of increasing due to accumulation of secretions.
LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP
The service life of LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP is influenced
by numerous factors. Thus, the composition of secretions, conscientious cleaning
and storage as well as other aspects can be of decisive significance.
The non-sterile LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP have an
average service life of up to 29 days if maintenance is performed regularly. After
that, the product must be replaced.
If LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP shows damage of any kind,
even if only minor, the product must be replaced immediately. LARYVOX
®
STYLE
TOP / LARYVOX
®
STYLE CAP that are presently not being used should be stored
in a dry environment, protected from dust, sunlight and heat.
IIX. CLEANING
LARYVOX
®
STYLE HME HighFlow filter cassettes
LARYVOX
®
STYLE HME HighFlow filter cassettes are single-patient products and
are intended for single use only. They must not be cleaned or disinfected, as use of
disinfectant can render the filter unusable.
LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP
Place LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP to soak in clear, lukewarm
water for approx. 2-3 minutes for cleaning, or use a mild, neutral-pH washing lotion.
Then rinse with plenty of clear water and allow to dry in air.
Do not use any sharp-edged cleaning utensils!
X. LEGAL NOTICES
The manufacturer Andreas Fahl Medizintechnik-Vertrieb GmbH will not
accept any liability for damages (especially not for functional deficiencies,
injuries, infections, and/or other complications or adverse events) caused
by unauthorised product alterations, by repairs not performed by the
manufacturer or by improper handling, care (cleaning/disinfection) and/or
storage of the products in violation of the provisions of these instructions for use.
This applies – to the extent permitted by law – both to damages to the products
themselves caused thereby and to any consequential damages caused thereby.
Product specifications described herein are subject to change without notice.