•  The user must check the 

Laryvox

®

 Hands-Free Valve

 for completeness, 

absence of damage and proper functioning of the product each time before 

inserting it again ➊!

• 

To prevent injury, it is advisable to first connect the speaking valve to the 

tracheostomy tube and then to insert the tracheostomy tube with attached 

speaking valve into the tracheostoma. Otherwise, excessive pressure could 

be  exerted  on  the  mucous  membranes  via  the  cannula  tube  when  the 

speaking valve is snapped into place.

•  You should adjust the speaking valve before attaching it and set the opening 

thickness of the valve disc by turning it ➋. The further the valve is opened, 

the more air can be inhaled. However, the exhalation pressure necessary to 

close the valve while speaking is then also greater. For normal respiration and 

normal physical activity, it is therefore recommended that you set the valve 

such that sufficient air can flow in, but without opening the valve completely. It 

is recommended that you open the valve to the maximum only for strenuous 

physical and sporting activities, in which your breathing becomes deeper.

•  After you have adjusted the valve disc to the desired opening thickness, 

insert the speaking valve together with the insertion ring, which can be seen 

on the small circumferential retaining bead, into the 22 mm adapter of the 

cannula ➌. 

•  If special base plates are used for adhesion (e.g. LARYVOX

®

 Tape), first 

fasten the base plate according to the instructions. 

•  Then carefully insert the 

LARYVOX

®

 Hands-Free Valve

 into the retaining 

ring of the ➍ base plate, so that a secure connection is established. 

•  If you are using a 

LARYVOX

®

 Hands-Free Valve

 for TrachiNaze

®

 and 

LARYVOX

®

 Hands-Free Valve

 Combi HME, now insert the additional filter 

➎ / ➏. 

•  The retaining bead ensures that the speaking valve will be seated stably. 

•  The strap (

LARYVOX

®

 Hands-Free Valve

 Low Resistance version) serves 

as a spacer to prevent restriction of breathing by clothing or the like. It is not 

suitable as a handle for positioning the speaking valve.

IX. Service Life

•  The service life of the 

LARYVOX

® 

Hands-Free Valve

 is influenced by many 

factors. Thus, the composition of secretions, conscientious cleaning and 

other aspects can be of decisive significance. 

•  The 

LARYVOX

®

 Hands-Free Valve

 has an expected life of twelve month 

when carefully treated. After that, the product must be replaced. 

•  A speaking valve showing damage of any kind, even if only minor, must be 

replaced immediately. 

•  Speaking valves that are presently not being used should be stored in a dry 

environment in a clean plastic container, protected from dust, sunlight and/

or heat.

Summary of Contents for 12.99.99.0005

Page 1: ...LARYVOX HANDS FREE VALVE GEBRAUCHSANWEISUNG INSTRUCTIONS FOR USE ...

Page 2: ...zeichnung Batch code P Gebrauchsanweisung beachten Consult instructions for use I Verwendbar bis Use by G CE Kennzeichnung CE marking 5 Einpatientenprodukt single patient use Q Inhaltsangabe in Stück Content in pieces m Bei beschädigter Verpackung nicht zu verwenden Do not use if package is damaged b Hersteller Manufacturer ...

Page 3: ... einzelnen Patienten auch zum Mehrfachgebrauch bestimmt Jeglicher Wiedereinsatz bei anderen Patienten und damit auch eine AufbereitungzumWiedereinsatzbeieinemanderenPatientenistunzulässig Reinigen Sie die LARYVOX Hands Free Valve Sprechventile nur mit vom Hersteller zugelassenen Reinigungsmitteln Eine Sterilisation des LARYVOX Hands Free Valve ist nicht möglich Das Erhitzen über 65 C das Auskochen...

Page 4: ...iden VII Produktbeschreibung Das LARYVOX Hands Free Valve Sprechventil gibt es in unterschiedlichen Ausführungen LARYVOX Hands Free Valve Low Resistance REF 21984 Es ist geringer Anblasdruck nötig um das Sprechventil zu schließen Der über dem Gehäuse verlaufende Bügel dient zur manuellen Feinabstimmung der Membraneinstellung und als Abstandshalter für Textilien LARYVOX Hands Free Valve Normal Low ...

Page 5: ...fiehlt sich nur bei anstrengenden körperlichen und sportlichen Aktivitäten bei der die Atmung verstärkt ist Nachdem die Ventilscheibe in der gewünschten Öffnungsstärke justiert wurde wird das Sprechventil mit dem Einsatzring der an dem kleinen umlaufenden Haltewulst zu erkennen ist in den 22 mm Adapter der Kanüle eingesetzt Bei der Verwendung von speziellen Basisplatten zum Aufkleben z B Laryvox T...

Page 6: ...Anleitung des Herstellers zu verwenden Zur Erleichterung der Reinigung empfehlen wir die Verwendung einer Kanülenreinigungsdose OPTIBRUSH CONT REF 31200 Spülen Sie anschließend das Sprechventil mehrmals gründlich mit klarem Wasser ab und trocknen Sie es mit einem sauberen weichen und fusselfreien Tuch Benutzen Sie keine scharfkantigen Reinigungsgeräte Alternativ ist auch eine Reinigung durch therm...

Page 7: ...dhabung begründet sind Insbesondere übernimmt die Andreas Fahl Medizintechnik Vertrieb GmbH keine Haftung für Schäden die durch Änderungen des Sprechventils oder durch Reparaturen entstehen wenn diese Änderungen oder Reparaturen nicht vom Hersteller selbst vorgenommen worden sind Dies gilt sowohl für hierdurch verursachte Schäden an den Sprechventilen selbst als auch für sämtliche hierdurch verurs...

Page 8: ...es are single patient products and are thus intended exclusively for use by one patient only They can be used several times by the same patient Re use of any kind in other patients and thus also reconditioning for re use in other patients is not allowed Clean the LARYVOX Hands Free speaking valves only with cleaning agents approved by the manufacturer The LARYVOX Hands Free Valve cannot be sterili...

Page 9: ...ifferent versions LARYVOX Hands Free Valve Low Resistance REF 21984 Only low blowing pressure is required to close the speaking valve The strap running across the housing can be used for manual fine tuning of the diaphragm setting and as a spacer for textiles LARYVOX Hands Free Valve Normal Low Resistance for TrachiNaze REF 21981 21980 Can be used in combination with TrachiNaze Plus filters The NO...

Page 10: ...g valve together with the insertion ring which can be seen on the small circumferential retaining bead into the 22 mm adapter of the cannula If special base plates are used for adhesion e g LARYVOX Tape first fasten the base plate according to the instructions Then carefully insert the LARYVOX Hands Free Valve into the retaining ring of the base plate so that a secure connection is established If ...

Page 11: ...mperature of not more than 65 C for this purpose Make sure that the temperature is kept constant use a thermometer to monitor the temperature and never under any circumstances expose the tubes to boiling water for disinfection This could seriously damage the speaking valve 2 Chemical Disinfection Instructions Disinfection should always be done if ordered by a physician due to specific health conce...

Page 12: ...thereby If the speaking valve is used beyond the period of use specified under Section IX and or if the speaking valve is used maintained cleaned disinfected or stored in non compliance with the instructions and specifications laid down in these instructions for use Andreas Fahl Medizintechnik Vetrieb GmbH will as far as legally permissible be free of any liability including liability for defects ...

Page 13: ...utilisées plusieurs fois ne doivent servir que pour un seul et même patient Toute réutilisation sur d autres patients est interdite il est donc également interdit de les retraiter en vue d une réutilisation sur un autre patient Ne nettoyez les valves de phonation LARYVOX Hands Free Valve qu avec les produits de nettoyage approuvés par le fabricant Une stérilisation de LARYVOX Hands Free Valve n es...

Page 14: ...révenir tout risque de blocage des voies respiratoires VII Description du produit La valve de phonation LARYVOX Hands Free Valve existe en plusieurs modèles LARYVOX Hands Free Valve Low Resistance RÉF 21984 Un léger souffle suffit pour refermer la valve de phonation L arceau au dessus du boîtier sert au réglage fin manuel de la membrane et à écarter les textiles LARYVOX Hands Free Valve Normal Low...

Page 15: ...ximale de la valve est recommandée uniquement pour les activités physiques et sportives exigeantes dans lesquelles la respiration s accélère Une fois le clapet ajusté au degré d ouverture souhaité insérez la valve de phonation et la bague d insertion reconnaissable au petit bourrelet de retenue rotatif dans l adaptateur de 22 mm de la canule Si vous utilisez des supports spéciaux adhésifs p ex Lar...

Page 16: ...aux instructions du fabricant Afin de faciliter le nettoyage nous recommandons d utiliser une boîte de nettoyage pour canules avec panier OPTIBRUSH CONT RÉF 31200 Ensuite rincez bien la valve de phonation plusieurs fois à l eau claire et séchez le avec un chiffon doux propre et non pelucheux Ne pas utiliser des outils de nettoyage pointus Un nettoyage par désinfection thermique à une température m...

Page 17: ...nik Vertrieb GmbH décline toute responsabilité en cas de dommages causés par les modifications des valves de phonation ou par des réparations si ces modifications ou réparations n ont pas été effectuées par le fabricant Ceci s applique autant aux dommages causés sur les valves de phonation qu à tous les dommages consécutifs en résultant Toute utilisation des valves de phonation dépassant la durée ...

Page 18: ...04 DOK GA Laryvox Hands Free Valve 01 2017 Andreas Fahl Medizintechnik Vertrieb GmbH August Horch Str 4a 51149 Köln Germany Phone 49 0 22 03 29 80 0 Fax 49 0 22 03 29 80 100 email vertrieb fahl de www fahl de b G ...

Reviews: