![Essilor MYOPIA EXPERT 700 User Manual Download Page 52](http://html.mh-extra.com/html/essilor/myopia-expert-700/myopia-expert-700_user-manual_2429035052.webp)
50
DESCRIPTION OF RESULTS
For each Biometry and Keratometry result, a dedicated section is present. In each section, the total result
is shown together with the standard deviation between the single results (if more than one) and any
warning or error marking (described in the following section).
The red arrows indicate to move the
instrument forwards towards the patient’s eye.
The centering condition is displayed in red
The blue arrows indicate to move the instrument backward away from the patient.
The centering condition is displayed in blue
The green signs indicate that the ideal focus has been reached. Press the joystick
button to start the automatic acquisition procedure.
The centering condition is displayed in green. Press the joystick
button to start the acquisition
The centering condition is displayed in orange. Center better
before pressing the joystick button to start the acquisition
The focusing or centering condition is very out of tolerance, and a better center and/or
distance alignment is required
2
1
Summary of Contents for MYOPIA EXPERT 700
Page 1: ...MYOPIA EXPERT 700 Biometer USER MANUAL Rev 1 07 04 2021...
Page 15: ...13 3 1 DEVICE SAMPLE LABELING...
Page 24: ...22...
Page 63: ...61 Delete Button Delete the selected row...
Page 79: ...77 Fig 43 It is possible to switch between ETDRS and Landolt C Visus simulation view...
Page 94: ...92 Fig 59...