112
For each service, the needed parameter is:
Remote Application Entity (AE) title
Remote IP address
Remote connection port
The
“
Local Application Entity title
”
is the identifier name through which the device presents itself to the
servers.
The
“
N-EVENT Report node port
”
is the port at which the device can receive N-EVENT REPORTS for
storage commitments (default is 115).
The connectivity to the defined server can be tested using the “C
-
ECHO” function activated by the related
“
Connection Test
”
button. The result of the connection test is shown by the green or red icon.
In order to configure the full DICOM workflow properly, it could be necessary to perform some operations
or configurations on the server’s side. In order to do this, contact the System Ad
ministrator.
The DICOM module of MYOPIA EXPERT 700 is described in detail in its
DICOM Conformance Statement
.
For more information, refer to your local distributor.