38
Fig. 11
Set the calibration tool supplied with the device (Fig. 12) in the special holes in the chinrest and press until
the tool is blocked on the device. Check that the calibration tool is perfectly aligned with the device. If the
calibration tool is positioned correctly, all the rings of the Placido disk should be seen reflected in the center
on the surface of the hemisphere (Fig. 13).
Fig. 12