Checks and Maintenance
78
Platinous User’s Manual
7.3.2 Water Filter Cleaning (ESL, ESX models only)
Washing of Strainer
Remove the cap, then pull out the filter from inside and wash it. At this time, deposits and foreign matter
sticking to the filter should be thoroughly removed with a brush. (Located in the water compartment, visible
when the access panel is removed).
FIGURE 7.46.
W
ASHING
STRAINER
FILTER
7.3.3 Humidifying Tray Cleaning
SHARP EDGES! Be careful of projections or sharp edges inside the
chamber.
For your safety, always wear gloves when working inside the chamber.
HOT ON THE INSIDE! Cool down the test area before cleaning the
humidifying tray.
During and shortly after operation above 55°C, the chamber is HOT and
HUMID on the inside (specimens, shelves, door gasket, test area walls.)
During operation, dirt and foreign matter stick to the humidifying tray and humidifying heater. To ensure
long lasting use, clean the tray and heater of dirt once a month. A good way to inhibit dirt from
accumulating is to drain the humidifying tray after every test.
!