Operation
68
Platinous User’s Manual
6.5 Useful Features
6.5.1 Auto/Manual Refrigeration Control
Refrigeration capacity is automatically controlled so as to reach and maintain the target temperature and
humidity. For normal tests, set refrigeration capacity to automatic control.
When refrigeration capacity is set to 100%, the refrigeration runs at
maximum capacity. In some cases it will not be possible to
maintain the target temperature. Check the capacity setting
BEFORE starting tests.
Specimens can be damaged if temperature cannot be maintained.
Note:
If trouble occurs during testing, set refrigeration capacity to manual control.
1. Useful if test samples create excess heat.
2. Useful if refrigeration needs to be turned off during operation.
Automatic Refrigeration Capacity Control
Procedure:
1. Set the main power switch in the ON position.
2. Press the POWER key to activate control power.
3. The following explains how to set refrigeration capacity in the constant mode. Press the CONSTANT
SETUP box.
Reference:
For an explanation on how to set refrigeration capacity in the program mode, see
“Program Create” in the SCP-220 User’s manual reference.
!
CAUTION
Summary of Contents for Platinous ESL Series
Page 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Page 2: ......
Page 6: ...iv Platinous User s Manual...
Page 22: ...Precautions in Usage 16 Platinous User s Manual...
Page 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Page 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Page 58: ...Installation 52 Platinous User s Manual...
Page 80: ...Operation 74 Platinous User s Manual...
Page 106: ...Troubleshooting 100 Platinous User s Manual...