Platinous User’s Manual
47
Installation
5.7.2 Water Supply Installation
Connection of the Water Supply Hose for the Humidifying Tray Water Circuit
Insert a 3/8” I.D. rubber hose, from a purified* water source, around the water supply hose nipple. The
water supply should have a pressure between 20 ~ 50 psi to ensure proper operation of the humidifying
circuit.
*The type of purifying system does not matter, but the purified water should be filtered (5 micron), have a conductivity of 0.2µS/cm to
10µS/cm, and a maximum of 2mg/L of free chlorine.
FIGURE 5.27.
C
ONNECTING
WATER
SUPPLY
HOSE
Water Hose
Water Port
Water Connector