Operation
64
Platinous User’s Manual
6.4 Test Startup/End
6.4.1 Test Startup
Before starting tests, make the preparations and settings described in Sections
to
“Constant Temperature (& Humidity) Setup” on page 62
Note:
Always wait at least 5 minutes after setting the main power switch in the OFF posi-
tion before reactivating the primary power supply to the chamber. Frequent
switching ON/OFF will greatly shorten refrigerator service-life.
Procedure:
1. Check control power is ON.
2. Get the Operation Mode Selection Screen. Press either the OPER./START key on the operating panel
or the chamber operating status box on the screen.
3. Start operation in the constant mode. Press the START button under Constant Mode.
4. Press the YES button on the confirmation message. The chamber will start up.
YES Button: Starts testing in the constant mode and returns the display to the original screen.
NO Button: Cancels the command and returns the display to the original screen. The
chamber’s operating status does not change.
6.4.2 Preventing Dew on Test Specimens
Reference:
When testing electronic components or similar specimens, it is necessary to keep
dew from forming on specimens. Dew forms when specimen surface temperature
is lower than the dew point of the air inside the chamber. To prevent dew forma-
tion, first warm specimens under temperature-only operations. Start temperature
and humidity test only after specimen surface temperature is the same as the
temperature inside the test area.
Summary of Contents for Platinous ESL Series
Page 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Page 2: ......
Page 6: ...iv Platinous User s Manual...
Page 22: ...Precautions in Usage 16 Platinous User s Manual...
Page 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Page 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Page 58: ...Installation 52 Platinous User s Manual...
Page 80: ...Operation 74 Platinous User s Manual...
Page 106: ...Troubleshooting 100 Platinous User s Manual...