Platinous User’s Manual
37
Installation
FIGURE 5.19.
I
NSTALLATION
SPACE
The front clearance* required for opening the door is as follows:
*Projections excluded
Chamber size is given below:
TABLE 5.9.
C
HAMBER
EXTERIOR
SIZE
TABLE 5.8.
R
EQUIRED
CLEARANCE
FROM
CHAMBER
FRONT
*
Type
ESL, ESU, ESX, ESZ
2
28 in. (70cm)
3
32 in. (80cm)
4
48 in. (120cm)
Type
ESL, ESU, ESX, ESZ
2
36.43”W x 47.08”H x 76.15”D
3
41.37”W x 55.01”H x 80.08”D
4
57.24”W x 58.66”H x 86.11”D
A
E
B
C
D
A: Min. 42” (106cm)*
B: Min. 6” (measured from top of motor cover)
C: See
below
D: Min. 24” (61cm)
E: Min. 24” (61cm)