Temperature Heat-up Rate
FIGURE 9.51.
T
EMPERATURE
HEAT
UP
RATE
Temperature-Humidity Con-
trol Range
The range in which temperature and humidity can be reproduced inside
the chamber. It is normally expressed as a graph with temperature as the
x-axis and humidity as the y-axis.
FIGURE 9.52.
T
EMPERATURE
-
HUMIDITY
C
ONTROL
R
ANGE
Temperature Pull-down Rate
The time required for the temperature to change from a defined tempera-
ture t1 to a defined lower temperature t2. This is simply the time required
for the drop in temperature and does not include the time it takes for the
temperature to stabilize.
FIGURE 9.53.
T
EMPERATURE
PULL
-
DOWN
RATE
Temperature Range
The range of temperatures that can be reproduced inside the chamber.
Temperature Uniformity
The difference in temperature between the geometric center of the
chamber and any arbitrary point in the effective area.
Test Chamber
A contained ambient constructed so as to obtain a desired temperature
and humidity.
Water-cooled Specification
(Water Cooled Condenser)
The condenser in the refrigeration circuit is cooled by water. Usually, a
cooling tower or similar unit is used to supply water and cool heat gener-
ated by the condenser. The water-cooled specification is an option with
Platinous Sterling Series Chambers.
t 2
t 1
T
e
m
per
at
ur
e
Time
Temperature
heat-up rate
20°C
40°C
80°C
60°C
0
20
40
60
80
100
100°C
85°C
90
98
%RH
10°C
15°C
45
t 1
t 2
Te
m
pe
ra
tu
re
T i m e
T e m p e ra tu re
p u l l-d o w n ra te
Summary of Contents for Platinous ESL Series
Page 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Page 2: ......
Page 6: ...iv Platinous User s Manual...
Page 22: ...Precautions in Usage 16 Platinous User s Manual...
Page 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Page 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Page 58: ...Installation 52 Platinous User s Manual...
Page 80: ...Operation 74 Platinous User s Manual...
Page 106: ...Troubleshooting 100 Platinous User s Manual...