Platinous User’s Manual
45
Installation
FIGURE 5.25.
H
UMIDITY
WATER
DRAIN
Note:
If a drainage pit is not located near enough to the chamber, run drainage into a
container. The humidifying trays use about 1.2 L per drainage cycle for chamber
types 2 and 3, and about 2.4 L for type 4 chambers. Whether running drainage
into a pit or a container, keep the end of the drain hose open to the atmosphere. If
submerged, water can backflow into the chamber.
3. Shut the water circuit compartment door. Attach the door in the opposite order in which it was detached.
Drain Connection
Drain Port
Drain Hose