Platinous User’s Manual
71
Operation
To power the lamp (equipment with 2A or more power)
Be sure to fit the line with a suitable capacity contactor and fuse.
FIGURE 6.42.
T
IME
SIGNAL
TERMINAL
(2A
OR
MORE
)
2. Turn the time signal ON for Steps 2 and 4 of the program.
Example 2: Following is an example 5-step program that charges specimens in steps 2 and 4.
FIGURE 6.43.
E
XAMPLE
PROGRAM
Procedure:
1. Connect the specimens to the time signal terminals and specimen power supply control terminals as
shown in the following diagrams.
Lam p
Fuse
Fuse
Power
supply
Time s ignal inlet
Target
temperature
(°C)
80
50
20
Time signal
ON
OFF
Step No.
1
2
3
4
5