36
490725-05
Appendix 5 : Technical data
POWER SUPPLY
Supply Voltage
Supply 230V (±10%) or 220V (-6%, +15%)
See rating plate. 50/60 Hz a.c.
For use with alternating current
Fuse Rating (panel)
2 x F10A, 250V (Part No. MUW218616)
Fuse Rating (plug)
13A
SAFETY CONDITIONS
Electric Shock Protection
Class 1
Pressure Relief Valve
Set pressure: 2.85 barg
STERILISING AND TEST DATA
Sterilising Time
3 minutes 20 seconds at 134/137°C
The sterilisation time can be increased to a maximum of
30 minutes by an Eschmann trained engineer.
Drying Time (approximate)
17 minutes
Total cycle time
11 mins (Empty, no drying)
Operating Pressure
303.5 kPa (abs) at 134°C, 331.1 kPa (abs) at 137°C
MAXIMUM LOADS
Maximum load per standard tray
1.5 kg
Maximum weight per half tray
0.75 kg
Maximum total load
4.5 kg
WATER AND RESERVOIR
Water Reservoir capacity/ Use per cycle
2500 ml/ 600ml (Proportion re-cycled)
Temperature of drained water/ flow rate
81
°
C/ 1.2 l/min
Heat transmitted by autoclave in one hour
720 kJ
DIMENSIONS AND WEIGHT
Depth
468 mm
Width
460 mm
Height
360 mm
Usable chamber space
190 x 110 x 288mm
Weight (empty)
26.5kg
Weight (full reservoirs and full load)
33.5kg
SHIPPING DIMENSIONS AND WEIGHT (includes unit, accessories and packing)
Depth
605mm
Width
655mm
Height
468mm
Weight
31.5kg
STANDARDS COMPLIANCE
BS EN ISO 15223-1
BS EN ISO 20417
BS EN 13060
BS EN ISO 14971
BS EN 61326
BS EN 62366
BS EN 13445
BS EN 61010
ENVIRONMENTAL REQUIREMENTS (contact Eschmann if required to operate outside these conditions)
Transport and Storage: (Note: Water must be drained)
Ambient temperature range
-10°C to +40°C
Relative humidity range
30-70%RH (Non-condensing)
Atmospheric pressure range
700 millibars to 1060 millibars
Operation:
Location
Indoor use only
Ambient temperature range
+5°C to +40°C