eschmann Little Sister SES 2010 User Manual Download Page 21

490725-05

 21

Viewing the Records Menu

1

SES 2010

09:00

SES 2010

09:00

Start Cycle

SES 2010

09:00

Cycle Type

Cycle No

 

The cycle has not yet started

From the Open Door, Close Door, or Cycle Start 
screens:

Press to view the Records menu. 

 

The Records menu 
contains two search 
criteria:

2

  Download Records - 

select from:

Successful Cycles 
All Cycles 
Failed Cycles

3

  Download Records - 

select from:

By Date 
All since last Download 
By Cycle Number

2

Settings 

SES 2010

09:00

Download Records

     

Successful Cycles

All Cycles

Failed Cycles

Download records

Use the up/down buttons to highlight 
your chosen setting in blue. 

                                                         
Press to accept and view the next 
screen.

Press to return 
without change.

3

Settings 

SES 2010

09:00

Download Records

     

By Date

All Since Last Down-

By Cycle Number

Download records

Use the up/down buttons to highlight 
your chosen setting in blue.

                                                         
Press to accept and view the next 
screen.

All Since Last Download

 will download all 

records produced since the last time it was used 
to download records.

Press to return 
without change.

Selecting All Since Last 
Download will begin the 
download immediately.

See below when selecting 
By Cycle Number and By 
Date.

13.2 Records Menu

By cycle number

1

SES 2010

09:00

Set FROM / TO Cycle Numbers

     

High Security

Current Cycle Number

00000

Set FROM / TO Cycle Numbers

Use the up/down buttons to highlight 
your chosen setting in blue.

Press to edit the selected cycle number.

 
Press to accept the cycle numbers 
shown and begin download.

Press to return 
without change.

The upper box (FROM) 
requires a cycle number.

The lower box (TO) 
always shows current 
cycle number, though this 
may be edited.

2

 Cycle Delay 

Keypad

SES 2010

09:00

2

3

4

5

6

7

8

9

0

1

00000

Insert the cycle number

Enter the cycle number in the keypad. Zero is 
not accepted.

Press 

OK

 to accept the cycle number entered.

Press to return 
without change.

Before beginning to download cycle records, insert a 
USB Memory Stick into the autoclave’s USB port  
(Refer to Fig. 1, page 4). 

Note:  Exit the Records menu before removing the 

USB device.

Summary of Contents for Little Sister SES 2010

Page 1: ...Powered by TCPDF www tcpdf org Ingenuity in infection control since 1830 User Manual Little Sister SES 2010 Non vacuum Autoclave...

Page 2: ...on 5 2 3 Unpacking 5 2 4 Autoclave positioning 5 2 5 Mains connection 5 2 6 Product packaging Markings 5 3 Controls 6 3 1 Touch screen menu operation 6 3 2 The Keypad 6 3 3 The Progress Screen 7 3 4 A...

Page 3: ...shooting In this case DO NOT switch off power until discharge of steam into the reservoir has stopped Do not attempt to open the chamber door until the unit has cooled 1 3 Limitations of use THIS IS A...

Page 4: ...on Bar CYCLE SELECTION Menu SETTINGS Menu RECORDS Menu 1 Chamber door 2 Door handle 3 Door catch 4 Touch screen display Fig 2 5 Mains on off switch 6 RS232 port 7 USB host port for USB memory stick 8...

Page 5: ...ote Recycle all packaging in accordance with local recycling practices Remove the outer packaging and all packaging pieces CAUTION Always move or lift the autoclave using at least two people Observe g...

Page 6: ...ever requires a User PIN Cycle Delay Keypad SES 2010 09 00 2 3 4 5 6 7 8 9 0 1 3 Controls 3 1 Touch screen menu operation Touch screen The autoclave must be switched on to use the touch screen The au...

Page 7: ...and maintenance of the autoclave Maintaining a record of training attendance including evidence of trainee understanding Refer to Section 13 Security to set up security PINS and choose a security lev...

Page 8: ...truments to be sterilised Load Cycle type Solid unwrapped instruments intended for immediate use N Maximum autoclave load 4 5kg Load trays using the supplied tray lifter All instruments MUST be thorou...

Page 9: ...essure kPa Indicates drying not selected WARNING During operation Do not remove or cover the water reservoir lid Fig 1 pg 4 Caution hot steam is vented Cycle progress Cycle Chart Cycle type Load type...

Page 10: ...door when OK button displayed WITHOUT DRYING WITH DRYING 1 Sterilisation complete SES 2010 09 00 Cycle Type Cycle No Temp C Pressure kPa 2 Positioning door for drying Take care The load chamber and do...

Page 11: ...cle failed There may be a delay whilst the chamber stabilises allowing the door to be opened 5 SES 2010 09 00 Cycle Type Cycle CYCLE COMPLETE Use Load Immediately This screen is only displayed if dryi...

Page 12: ...ve The red drain tube clip is now holding the water back Open the drain tube clip to drain the water 9 2 Switch off Switch off the autoclave at the mains switch Fig 1 pg 4 when not in use Leave it wit...

Page 13: ...nd decontamination when disposing of the door seal 3 Update the Settings Menu counter for the door seal Updating the Settings Menu It is important to update the door seal counter in the Maintenance Me...

Page 14: ...available whilst the autoclave is depressurising in preparation for the door being opened This may take up to two minutes at which point one of the following Warning screens will be displayed Refer to...

Page 15: ...schmann 1 Note the Advisory Code When it appears press OK to acknowledge the warning and clear the screen 2 Refer to Appendix 4 and complete the Actions against the relevant Advisory Code 3 Run the cy...

Page 16: ...Start screens Press to view the Cycle Selection menu The 134 N cycle is the factory default for each new cycle The current cycle type is shown on the Start screen 2 Cycle Select SES 2010 09 00 Cycle S...

Page 17: ...s to accept the delay time shown and go to 5 below or Press to enter a new delay time Press to return without change 4 Cycle Delay Keypad SES 2010 09 00 2 3 4 5 6 7 8 9 0 1 00 00 Insert the Delay Time...

Page 18: ...the 24 hour clock Press to return without change 3 Date Time Date entry SES 2010 09 00 Date Time 13 00 01 10 2019 Daylight Saving Set the autoclave s date and time Press OK to accept Press to return...

Page 19: ...larms Cycle Alarms Cycle Keypress Note that alarms are always sounded The selection highlighted in blue is the current selection PIN management SES 2010 09 00 PIN Management 0000 3 2 1 Manager Setting...

Page 20: ...highlighted setting Press OK to accept Press to return without change Replace the filter and or seal before resetting the counter s When you reset an item on this screen the item is marked with an arr...

Page 21: ...ince Last Download will download all records produced since the last time it was used to download records Press to return without change Selecting All Since Last Download will begin the download immed...

Page 22: ...atted memory stick is full memory stick was unplugged search criteria returned no data Correct and try again By date 1 SES 2010 09 00 Set FROM TO Dates High Security Today s Date 01 30 __ __ __ Set FR...

Page 23: ...codes will allow you to know Who started a cycle High security level Who stopped a cycle All security levels Who unlocked the door at the end of a cycle All security levels Who cleared Service screens...

Page 24: ...a Manager PIN 0000 is an unused PIN is a used PIN 4 Pin Entry SES 2010 09 00 PIN Configuration User 2 0002 2 Cycle Delay Keypad SES 2010 09 00 2 3 4 5 6 7 8 9 0 1 0002 Select to Edit PIN or status lev...

Page 25: ...ue area of the screen Press to accept Press to return without change 3 SES 2010 09 00 Security Level Low Security Medium Security High Security Select required security level Use the up down buttons t...

Page 26: ...older on your PC desktop To create the back up folder right click on the desktop and select New then Folder Type in the name of the new folder We suggest SES 2010 If you are backing up more than one u...

Page 27: ...elow 2 Click the arrow to select your printer 3 Click to print the file Printing options toggle to display all the data in the cycle or only major events toggle to select double sided or single sided...

Page 28: ...hh mm ss dd mm yy ERR error STS status at cycle end PCV pressure change value Service and certification tags SVC service expired date cycle count of service expiry CRT certification expired date of ce...

Page 29: ...ator must only be used and stored as detailed in the Instructions For Use supplied with the Device Failure to do so could lead to failure of the Device and dangerously misleading results This device s...

Page 30: ...pendix 5 Maximum Loads If in doubt about loads and processing details contact Eschmann Unwrapped solid items only The autoclave is ONLY suitable for unwrapped solid items Refer to Section 1 3 Limitati...

Page 31: ...ogenic material from instruments used in the Health sector Eschmann are able to advise on how to recycle and dispose of the product correctly Example 1 6 x HFiT10 Cassettes 60 Instruments Example 2 2...

Page 32: ...al legislation WARNING Should the autoclave ever leak disconnect from the mains electrical supply drain the reservoir and call Eschmann Autoclave fuses Fuse x 2 Fig 1 pg 4 Refer to Appendix 5 Technica...

Page 33: ...ion Power autoclave off on Run cycle again 25 Not Enough Water in Reservoir Fill with fresh water Run cycle again 26 Too Much Water in Reservoir Empty waste water Run cycle again 33 Door Not Closed Po...

Page 34: ...cycle again 175 Pressure Not Equalised Power autoclave off on Run cycle again 176 Safety Relay Fail Visually check door seal Clean mating face Run cycle again or Leave autoclave to cool Safety relay...

Page 35: ...Mismatch Contact Eschmann ERROR 18 Pressure Switch Failed Contact Eschmann 177 Failed To Update Calibration Record Contact Eschmann 178 Calibration Invalid Contact Eschmann 179 No valid configuration...

Page 36: ...4 5 kg WATER AND RESERVOIR Water Reservoir capacity Use per cycle 2500 ml 600ml Proportion re cycled Temperature of drained water flow rate 81 C 1 2 l min Heat transmitted by autoclave in one hour 72...

Page 37: ...et Pressure 2 85 bar Hydrostatic Test Pressure 4 15 bar Pollution degree 2 SAFETY FEATURES Eschmann autoclaves are designed to be safe and effective No changes should be made except by an Eschmann Eng...

Page 38: ...lid load Small porous loads Full porous load Narrow Lumen Dryness solid load Dryness porous load Non condensable gases Load chart Cycle Data Sterilisation Temperature 134 137 C Sterilisation Pressure...

Page 39: ...nd cycle codes Absolute Pressure Cycle stages 1 to 5 Detailed N type cycle information Progress Indicator Cycle stage Stage Detail 1 2 1 Filling and 2 heating to sterilisation temperature 3 Sterilisin...

Page 40: ...FiT Cassettes HFiT 10 Examination Stainless Steel Cassette 13 5cm x 18cm 8722812 Stainless Steel Blue 8722813 Stainless Steel Red 8722814 Stainless Steel Green 8722815 Stainless Steel Yellow 8722816 S...

Page 41: ...Powered by TCPDF www tcpdf org 490725 05 41...

Page 42: ...n this publication may be subject to change without prior notice Service The Autoclave must only be serviced by an Eschmann or Eschmann trained engineer We also recommend a program of ongoing service...

Reviews: