34
490725-05
Advisory
screens
Advisory
code
Description
Action
43
Missed First Process Assessment
Number
Power autoclave off / on. Run cycle again.
44
Process Assessment Sequence Error
Power autoclave off / on. Run cycle again.
45
Error Unexpected End of Process
evaluation List
Power autoclave off / on. Run cycle again.
47
Error Pressure and Temperature
Disagree
Visually check door seal. Clean mating face.
Run cycle again.
48
Error Process Assessment Failed
Power autoclave off / on. Run cycle again.
49
Drying Failed
Power autoclave off / on, run cycle again.
Run without drying.
155
Communications Failure
Power autoclave off / on. Run cycle again.
156
Communications Request Failure
Power autoclave off / on first, then press OK to
clear the SYSTEM STOPPED screen. Run cycle
again.
157
Communications Remove Failure
Power autoclave off / on first, then press OK to
clear the SYSTEM STOPPED screen. Run cycle
again.
160
System Synchronisation Failure
Power autoclave off / on. Run cycle again.
175
Pressure Not Equalised
Power autoclave off / on. Run cycle again.
176
Safety Relay Fail
Visually check door seal. Clean mating face.
Run cycle again
or
Leave autoclave to cool. Safety relay resets
automatically. Run cycle again.
201
Enclosure Temperature Too High
Leave autoclave to cool. Run cycle again.
202
Process Assessment Temperature Too
High
Power autoclave off / on. Run cycle again.
203
Process Assessment Temperature Too
Low
Power autoclave off / on. Run cycle again.
204
Process Assessment Pressure Too
High
Power autoclave off / on. Run cycle again.
205
Process Assessment Pressure Too Low Power autoclave off / on. Run cycle again.
209
Process Assessment Exceeded Time
Limit
Power autoclave off / on. Run cycle again.
210
Process Assessment Sterilisation
Temperature Too Low
Power autoclave off / on. Run cycle again.
211
Process Assessment Sterilisation
Temperature Too High
Power autoclave off / on. Run cycle again.
217
Data Corrupt
Power autoclave off / on first, then press OK to
clear the SYSTEM STOPPED screen. Run cycle
again.
223
Touch Screen Continuously Pressed
Power autoclave off / on. If persistent contact
Eschmann.
226
Cycle Check Failed
Power autoclave off / on. Run cycle again.