28
490725-05
Cycle output
The cycle output (opposite) can provide the following
information:
[SRN]
autoclave serial number
[CYC]
cycle type (e.g 134°N) and
cycle count (e.g. 00022)
[STD]
start time and date
(hh:mm:ss, dd\mm\yy)
Sterilisation information (time, pressure
and temperature every 60 seconds during
sterilisation)
[TPP]
time of pressure point
[TSS]
time of sterilisation start
[p]
print sterilisation line
(“all data” viewing only)
[TSE]
time of sterilisation end
[TMX]
maximum sterilisation temp
[TMN]
minimum sterilisation temp
[TDS]
time drying start
[TDE]
time drying end
[PTD]
time cycle finished processing
(“all data” viewing only)
(hh:mm:ss, dd\mm\yy)
[ETD]
end time and date
(hh:mm:ss, dd\mm\yy)
[ERR]
error
[STS]
status at cycle end
[PCV]
pressure change value
Service and certification tags
[SVC]
service expired
(date/cycle count of service expiry)
[CRT]
certification expired
(date of certification)
Security log additions
(displays the number of the PIN: 1-12, 0 = not entered)
[DCB]
drying cancelled by
[CSB]
cycle started by
[CAB]
cycle aborted by
[LRB]
load retrieved by
[LRT]
(pin not
displayed)
time user retreived load
(“all data” viewing only)
(hh:mm:ss, dd\mm\yy)
Interpreting the output
To indicate a successful cycle, [STS] displays
“Cycle Complete” for 1
34°C N cycles, or “Calibration
Complete” for calibration cycles
.
If any cycle parameters are incorrect, [STS] displays
***FAIL***, followed by an Advisory Code. Refer to
Appendix 4.
For 1
34°C N cycles,
you can check that:
Temperature
[TMX] and [TMN] is between 134-137
°C
Time
between [TSS] and [TSE] is >3min.20secs
Pressure
@134°C = 303.5kPa, @137°C = 331.1kPa