14
490725-05
SES 2010
09:00
CRITICAL
ERROR
CYCLE
STOPPED
Load Not
Sterilised
POWER
FAILURE
Load Not
Sterilised
Advisory Code:
1
Displayed if the autoclave experiences a power
interruption. Note: Door lock will stay in (Or change to)
the locked state, upon power failure.
Press to display the details required for
contacting Eschmann.
1. When it appears. press OK to
acknowledge the warning and clear the
screen.
2. Check mains supply, mains connections,
autoclave fuses and mains cable fuse.
3. Run the cycle again. If the warning
persists, contact Eschmann.
Eschmann use only (PIN protected).
POWER FAILURE
Advisory screens may be displayed at any time to
protect the autoclave.
When an Advisory screen is first displayed some
touch screen buttons are not available whilst the
autoclave is depressurising in preparation for the
door being opened.
This may take up to two minutes, at which point one
of the following Warning screens will be displayed.
Refer to Appendix 4. Trouble shooting, Table 1.
SES 2010
09:00
USER
STOPPED CYCLE
Load Not
Sterilised
Advisory Code:
2
Displayed when the STOP button
is pressed to
stop a cycle.
Press to display the details required for
contacting Eschmann.
1. When it appears, press the OK button to
acknowledge the warning and clear the
screen.
Eschmann use only (PIN protected).
USER STOPPED CYCLE
WARNING!
If an advisory screen is displayed, the load must be treated as NON-STERILE.
Sterilise the load by running the cycle again.
Do not attempt to solve the problem using tools or by tampering with the autoclave.
CAUTION
The water, chamber and load could be very hot
and potentially non-sterile.
After clearing any warning, always allow the chamber
to dry before starting another cycle.
11 Advisory screens