490725-05
35
Advisory
screens
Advisory
code
Description
Action
228
Start Up Test Failed
Power autoclave off / on first, then press OK to
clear the SYSTEM STOPPED screen. Run cycle
again.
231
Calculated Temperature Too High
Power autoclave off / on. Run cycle again.
232
Calculated Temperature Too Low
Power autoclave off / on. Run cycle again.
233
Calculated Temperature Different To
Actual Temperature
Power autoclave off / on. Run cycle again.
909
Chamber Temperature Sensor
Power autoclave off / on. Run cycle again.
910
Chamber Pressure Sensor Fault
Power autoclave off / on. Run cycle again.
913
Enclosure Temperature Sensor Fault
Power autoclave off / on. Run cycle again.
916
Unknown Sensor Fault
Power autoclave off / on. Run cycle again.
917
Configuration Mismatch
Contact Eschmann.
ERROR
18
Pressure Switch Failed
Contact Eschmann.
177
Failed To Update Calibration Record
Contact Eschmann.
178
Calibration Invalid
Contact Eschmann.
179
No valid configuration
Contact Eschmann.
206
Safety Valve Failure
Contact Eschmann.
208
Safety Valve Failure
Contact Eschmann.
901
Failed To Allocate Timer
Contact Eschmann.
902
Failed to Allocate PWM Channel
Contact Eschmann.
903
Unrecognised command
Contact Eschmann.
905
Unable to find unit’s configuration
Contact Eschmann.
917
Configuration mismatch
Contact Eschmann.
918
Control system set then cleared arbiter
at end of cycle
Contact Eschmann.
919
Wrong unit type
Contact Eschmann.
920
Unknown unit type
Contact Eschmann.
921
CO not fully configured
Contact Eschmann.
922
Versions conflict
Contact Eschmann.
980
Run out of data blocks
Contact Eschmann.
981
Run out of cycle blocks
Contact Eschmann.
982
Not in list of cycle blocks
Contact Eschmann.
983
Cycle block not set up
Contact Eschmann.