![Esaote MyLabOmega Getting Started Download Page 54](http://html1.mh-extra.com/html/esaote/mylabomega/mylabomega_getting-started_2425221054.webp)
P R E P A R I N G T H E S Y S T E M
MyLab
- G E T T I N G S T A R T E D
4 - 2
Connecting Peripherals
Peripherals, that have been ordered simultaneously with the
MyLab
, are
usually already mounted and connected. The first mounting and connecting
will usually be performed by an Esaote technician.
NOTE
Esaote suggests to contact its Service representative to install any auxiliary
device.
Before installing the peripheral devices, make sure that the system is switched
off and unplug the power cable from the mains outlet. Brake the wheel to fix
the system.
Powering sockets for peripheral devices are placed at the rear on the left;
peripheral connections are placed at the rear, on the right. The network
connector is placed at the rear in the bottom central position.
How to connect peripheral devices:
1. Ensure that the
MyLab
is switched off (complete shut down
not stand-by or other conditions).
2. Connect the peripheral device to the
MyLab
.
3. Switch the peripheral device on, making sure that the device
is not in stand-by condition.
4. Switch the
MyLab
on by pressing the Power ON button.
NOTE
Always observe the instructions given in the manual of the peripheral/
auxiliary device.
NOTE
Contact Esaote personnel for recommended USB printers and for safe and
proper installation.
NOTE
Not all the external monitors are compatible with
MyLab
. Please contact
your Service representative to select an external monitor that can be
managed by the system.
Safety Concept
MyLab
is equipped with an insulation transformer to provide required
separation from AC mains for both the system and the auxiliary devices. Two
plugs for connecting auxiliary devices are located in the back of the system
and are accessible opening the rear door.
Additional equipment connected to
MyLab
must comply with respective IEC
or ISO standards (e.g. IEC 60950 for data processing equipment).
Summary of Contents for MyLabOmega
Page 1: ...Rev 04 July 2019 MyLabOmega GETTING STARTED 350035100...
Page 2: ...MyLab G E T T I N G S T A R T E D ii...
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity...
Page 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED...
Page 8: ...MyLab G E T T I N G S T A R T E D viii...
Page 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv...
Page 26: ...I N T R O D U C T I O N MyLab G E T T I N G S T A R T E D 1 12...
Page 52: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 18...
Page 62: ...P R E P A R I N G T H E S Y S T E M MyLab G E T T I N G S T A R T E D 4 10...